Description
Insulation Resistance Range :??0.1M?? ~ 10G??
Selectable Test Voltages(Low Voltage for Telecom) : 50 / 100 / 125 / 250 / 500 / 1000V
LCD Type :??Large Type 2 x 16 Characters
Continuity Measurement :??0.01 to 1999
Automatic AC / DC Voltage Test At Start-Up
AC Voltage Measurement :??0 to 700V
DC Voltage Measurement :??0 to 950V
Battery Load Current :??About 300mA
Battery Test at switch ON / RESET Lead Resistance Null Facility
Auto Off approx. 5 minutes
Operating Temperature :??1 ??C to 55 ??C not in full sun
Storage Temperature : -20 ??C to 70 ??C ; Relative Humidity max. 80% R.H.
Open Current :??about 19.15mA
Consumption Current :??about 193.5mA.
Power Supply : 8 x1.5V Alkaline batteries.
Dimension :??175(L) x 85(W) x 75(H) mm
Weight : Approx. 655gms.
Auto Discharge Function on all test & all ranges.
Gas Arrester Test
MOV / Protection Devices Test
Preselectable Measurement Time : Auto Stop on completion of Test
Auto Stop & Hold Function.
Ener Save ???
Can be calibrated in All calibration laboratories
Buzzer ON/OFF by key : Always ON
Nominal Voltage @ 1mA on all Insulation Ranges
Continuity Short Circuit Current >220mA (225mA typical)
Continuity Open Circuit Voltage of 5V DC
Smart Hold & Stop on Voltmeter AC / DC
Test ON/OFF
Fuse : fast HBC 600V / 500mA